Benchtop
N<sub>2</sub>O/CO Analyzer

N2O/CO Analyzer


Performance

Concentration Range

N2O: 1 - 4000 ppbv
CO: 1 - 4000 ppbv

Data Rate

0.01-20 Hz (external pump may be required)

Repeatability/Precision

N2O: 50 pptv (3 minute)
N2O: 0.3 ppbv (1 second)

CO: 50 pptv (3 minute)
CO: 0.3 ppbv (1 second)

Size & Weight

Dimensions

10" H x 38" W x 14" D
(includes embedded video monitor, keyboard, mouse)

Weight

80 lbs (36 kg)

Operating Requirements

Ambient Humidity

<98% RH Non-Condensing

Operating Temperature

5-45 °C

Power Requirements

115/230 VAC; 50/60 Hz
150 watts

Sample Temperature

0-50 °C

Configuration

Data Storage

Internal Hard Drive

Display

12.1" Color TFT

Inlet/Outlet Fittings

3/8", 1/2" Swagelok®
3/8", 1/4" Swagelok®

Outputs

Digital (RS232), Analog, Ethernet

Ordering Information


Price: $89,950 US list price; prices will vary by region
Model Number: 908-0014

Options

Multiport Inlet unit
Enables automated control of 16 inlet ports
Dry-Scroll Vacuum Pump (DSVP)
Enables measurement rates >1 Hz

Accessories

Multiport Inlet Unit
LGR's MIU is a separate box with 16 input ports and 1 output port  connected to the Analyzer. The MIU inputs are fully controlled by software on the LGR Gas Analyzer so that any port may be selected at any time. Once programmed, the sequence of measurements is fully automated.

Download data sheet